自动组织脱水机
仪器介绍
自动组织脱水机用于将组织样本进行脱水处理,并用石蜡渗透。收获的组织样本可进行下一步石蜡包埋和切片实验。它通过合适的搅拌方式,在不降低处理质量的情况下,减少了组织处理时间,降低了组织损伤的风险。同时,封闭的一体化结构可以减少用户接触试剂的情况。
参数
Specimen capacity | The maximum specimen processing capacity is ≥300 embedding cassettes. |
Reagent stations | ≥16 reagent stations (including 6 dehydration reagent bottles; 3 transparent reagent bottles; 2 fixative reagent bottles and 2 replacement reagent bottles; 3 flushing stations, capacity 5L). |
Wax tank | Metal wax tank, 3 wax tanks and 1 waste wax tank, capacity: ≥5.6L; temperature range: 45°C-65°C; with dual temperature detection function, real-time monitoring of paraffin level through temperature comparison. |
Reagent preheating function | With reagent preheating function, the dehydrating agent and transparent reagent will be dehydrated faster after preheating, and the effect will be more uniform; reagent storage temperature: room temperature ~ 35 ℃. |
Alcohol specific gravity monitoring function | With alcohol specific gravity monitoring function, the quality of reagents can be continuously tracked to ensure high-quality dehydration; reagent rotation can be automatically completed inside the machine without interrupting the dehydration process, avoiding direct contact between operators and harmful gases. |
Automatic replenishment function | It has the function of automatic replenishment of less liquid and the function of reagent transferring, which can ensure the safety of samples in all aspects. |
Backup battery | Equipped with a backup battery, the standby time is ≥ 4 hours, the short program can be run directly, and the long program can be ensured the safety of the sample. |
Filtration system | The reaction cylinder has a downward pumping function, which can effectively control the emission of harmful gases; a double filter system of potassium permanganate and activated carbon is used to ensure the safety of the operator. |
Xylene-free programs | Xylene-free programs can be run without heating up (to 85°C), and will not affect subsequent IHC studies. |
End time | The extension end time can be programmed and set, even unlimited extension |