气相色谱质谱联用仪

型号:

Agilent 8890/5977C GC/MSD

地点:

132, 1F, E4

仪器介绍

气相色谱法(GC)是一种适用于分离和分析挥发性、半挥发性且热稳定化合物的色谱技术。Agilent 8890/5977C GC/MSD 系统是一套高性能的气相色谱-单四极杆质谱联用系统。其核心的电子压力控制(EPC) 技术确保了全系统出色的压力和流量精度,为方法的重现性和可靠性奠定了基础。微板流路控制(CFT) 技术不仅实现了精确的流量分配,更支持检测器分流、色谱柱反吹和多维色谱等高级应用,例如可选配将色谱流出物同时送至质谱检测器(MSD,用于定性鉴定)和火焰离子化检测器(FID,用于高精度定量),从而通过多种数据源的互补全面增强分析能力。气质联用(GC-MS)技术广泛应用于环境监测、食品安全、石化分析及法医学等领域,是各类实验室进行常规靶向分析的主力机型,以可靠和高效著称。

参数

8890 GC System
Column ovena) Operating temperature: 4°C to 450°C above room temperature.
b) Temperature setting: Step size 1°C, program heating interval 0.1°C.
c) Heating rate: programmable heating, 120°C /min; Programmable cooling, can be reduced from 400°C to 50°C within 4 minutes.
d) Program heating: 20 stages, 21 temperature platforms.
e) Stability: ≤ 0.01°C /1°C (with a variation range of 0.01°C for every 1°C).
f) Column replacement without venting.
Split/non split injection porta) The maximum temperature is 400°C.
b) Pressure range: 0-100 psi, electronic pressure control accuracy: 0.001 psi
GC-FIDa) Electronic pressure/flow control (electronic pressure control).
b) Equipped with automatic shutdown detection function.
c) Sampling rate of 1000 Hz.
d) Minimum detection limit: ≤ 1.2 pg carbon/second
Liquid autosamplera) Number of positions for sample vials (2 mL-vial): 50 positions, 3 self-cleaning positions.
b) Injection volume range: between 0.1-10 µL
Mass ResolutionLinear mode ≥ 1,200 for Cytochrome C (m/z 12,361) FWHM; Reflect mode ≥ 50,000 for Somatostatin 28 (m/z 3,147.5) FWHM; MS/MS Fragment Resolution,≥ 1,500 for [Glu1]-Fib fragment m/z 175, ≥ 5,000 for [Glu1]-Fib fragment m/z 684, ≥ 7,500 for [Glu1]-Fib fragment m/z 1,056, ≥ 10,000 for [Glu1]-Fib fragment m/z 1,441
MS SensitivityLinear mode, 500 fmol BSA (m/z 66,000) at S/N ≥ 100:1 shown with 1,000 laser shots;Reflect mode , 250 amol [Glu1]-Fibrinopeptide B (m/z 1,570.7) at S/N ≥ 200:1 shown with 2,000 laser shots; MS/MS mode: 250 amol [Glu1]-Fibrinopeptide B fragment 1,056 Da at S/N ≥ 30:1 shown with 1,000 laser shots
5977C MSD System
EI sourceIonization energy 5-241.5 eV
Ion source temperature350°C
Quadrupole mass analyzera) A quartz gold-plated conjugate hyperbolic quadrupole
b) Independent temperature control
c) Temperature control range: 106-200°C
d) Maintenance free for life
Signal-to-noise ratioEI full scan ≥ 5000:1 s/n (1pg OFN, scanning range 50-300 amu)
Scanning speedup to 20000 amu/s
Mass range0.6~1091 Da
Mass stability<0.10 amu/48h
Dynamic range≥ 106

手册

Agilent 8890/5977C GC/MSD_Brochure_EN