Single‑Quadrupole LC‑MS System

Model:

Agilent 1290 Infinity Ⅲ- Pro iQ

Location:

132, 1F, E4

Equipment Introduction

Agilent 1290 Infinity Ⅲ‑Pro iQ -single quadrupole LC-MS system. Its LC module offers a maximum pressure tolerance of 1300 bar and low extra-column dead volume, enabling efficient and rapid separation. The integrated Pro iQ mass detector features fast scanning, a wide mass range and high sensitivity. With a compact footprint, it can be stacked with the UHPLC system and supports ESI and APCI ionization modes. Seamlessly interfaced with the matched LC platform and OpenLab CDS workstation, the instrument streamlines qualitative and quantitative workflows. Equipped with intelligent status alert and auto-tune functions, it delivers excellent system stability and easy maintenance. It is mainly used for qualitative and quantitative analysis of small molecules, impurity profiling, peptide molecular weight determination, and screening of environmental contaminants. Widely applied in drug development, metabolomics, proteomics and environmental analysis, the instrument is open for shared access by researchers.

Specification

Agilent 1290 Infinity III UHPLC
Quaternary PumpFlow precision: RSD ≤ 0.07 %
Flow rate range: 0.001 - 5 mL/min, minimum setting increment: ≤ 0.001 mL/min
Operating pressure range: 0 1300 bar
Gradient composition precision: RSD < 0.15 %
Integrated degasser: 4 channels
AutosamplerInjection volume range: 0.1 - 20 μL, minimum increment: ≤ 0.1 μL
Injection precision: RSD < 0.15 %
Operating pressure range: 0 - 1300 bar
Sample capacity: 2 drawers, 200 positions for 2 mL vials; microplate injection supported
Sample compartment temperature control: 4 ℃ - 40 ℃
Column ovenTemperature range: 4 - 110 °C
Temperature control stability: ≤ ±0.03 °C
Temperature control accuracy: ≤ ±0.5 °C
Diode Array Detector (DAD)Number of photodiodes: 1024
signal channels: 8
Maximum sampling rate: 240 Hz
Short term noise at 230 nm: ≤ ± 3 × 10⁻⁶ AU
Drift at 230 nm: ≤ 0.5 × 10⁻³ AU/hr
Wavelength range: 190 - 640 nm
Programmable slit width: 1, 2, 4, 8 nm

Single Quadrupole Mass Spectrometer
Ion sourceESI and APCI ion sources; cleaning and replacement of ion transfer unit without venting vacuum
Mass scan rangem/z 2 - 1600
Maximum scan speed15,000 amu/s
Sensitivity• ESI positive mode: Column infused 1 pg reserpine, S/N ≥200:1 at m/z 609.3 in SIM mode
• APCI positive mode: Column infused 1 pg reserpine, S/N ≥70:1 at m/z 609.3 in SIM mode
Positive/negative polarity switching time≤15 ms
Dynamic range≥ 6×10⁶
Minimum SIM dwell time1 ms
Resolution mode4 types, supporting 0.4 Da high resolution mode
Mass accuracy≤ ±0.1 Da
Mass stability≤0.1 Da /24 hrs

Brochure

Agilent InfinityLab Pro iQ_Brochure_CN_1

Agilent InfinityLab Pro iQ_Brochure_CN_2