Analytical Ultracentrifuge

Model:

Optima AUC

Location:

The hallway of 126-133

Equipment Introduction

Optima AUC is a general method for the qualitative analysis of macromolecules in solution. The Optima AUC combines a centrifuge capable of settling particulate matter with an optical module capable of detecting settling over time. Analytical ultracentrifugation techniques provide insight into sample molecular weight, shape, structure, interactions, and heterogeneity.

Specification

Maximun speed60,000 rpm
TimerUp to 999 hrs 59 min, including Hold runs
Temperature7 to 40°C in 1°C increments at >25°C to 30°C ambient temperature; 4 to 40°C in 1°C increments at <25°C ambient temperature
Analytical acceleration5 to 400 rpm per second above 500 RPM; 400 is maximum acceleration
Analytical deceleration5 to 400 rpm per second above 500 RPM; 400 is maximum deceleration
RotorsAN-60 Ti with 4 sample tanks
maximun speed: 60,000 rpm
AN-50 Ti with 8 sample tanks
maximun speed: 50,000 rpm
Binding/dissociation constants10-3 - 10-8 M
Sample cells6 and 2
Scanning UV/VIS absorbance optical systemWavelength range: 190 to 800 nm;
Photometric display range: 0 to 3 absorption units;
Scan steps: Radial scan mode - up to 1,500 data points per scan; radial increment step size 0.001 cm minimum, 0.01 cm maximum Wavelength scan mode - wavelength increment step size 1.0 nm, from 190 nm to 800 nm;
Scanning interval: 1 nm;
Data rate: One intensity measurement every 20 ms.
Interference optical systemWavelength: 660 nm;
Camera resolution: 2048 x 1088 pixels;
Laser: 120 mW Diode;
Scan rate: Approximately every 5 seconds;
Interferometer precision: Approximately ± 0.008 fringe.