Fluorescence Spectrophotometer

Model:

F-7100

Location:

132, 1F, E4

Equipment Introduction

F-7100, the compact system capable of performing "Best-in-Class" Analytical Signal-to-Noise and ultra high-speed (60,000 nm/min) at the highest level of its class. The application of fluorophotometry has been expanded to various fields such as industrial materials, in areas such as organic electroluminescence and liquid crystals; environment-related areas, such as water quality analysis; pharmaceutical manufacturing, such as the synthesis and development of a fluorescence reagent; and to biotechnology-related areas, such as intracellular calcium concentration measurement.

Specification

Sensitivity (Raman light of water)Noise: Background S/N 20,000 or above (EX 350nm, Slit 10 nm, Response 4 s);
Noise: Peak S/N 1,200 or above (EX 350nm, Slit 5 nm, Response 2 s)
Minimum sample volume0.6 mL (in use of standard 10 mm rectangular cell)
Photometric principleMonochromatic light monitoring ratio calculation
Light source150 W xenon lamp, self-deozonating lamp house
MonochromatorStigmatic concave diffraction grating: 900 lines/mm, F2.2;
Brazed wavelength: Excitation side 300 nm, emission side 400 nm
Measuring wavelength range (on both EX and EM)200 to 750 nm, and zero-order light (Expandable up to 900 nm with optional detector)
BandpassExcitation side: 1, 2.5, 5, 10, 20 nm;
Emission side: 1, 2.5, 5, 10, 20 nm
Resolution1.0 nm (at 546.1 nm)
Wavelength accuracy±1 nm
Wavelength scan speed30, 60, 240, 1,200, 2,400, 12,000, 30,000, 60,000 nm/min
Wavelength drive speed60,000 nm/min
3D measurement time3 min (EX 200 to 750 nm, Sampling interval 10 nm, EM 200 to 750 nm, Sampling interval 10 nm)
ResponseResponse from 0 to 98 %: 0.002, 0.004, 0.01, 0.05, 0.1, 0.5, 2, 4 s
Photometric value range-9999 to 9999

Brochure

F-7100_Brochure_EN