Multimode Microplate Reader

Model:

Varioskan LUX

Location:

127, 1F, E4

Equipment Introduction

Designed for bioscience researchers with a wide variety of needs, the Thermo Scientific™ Varioskan™ LUX multimode microplate reader comes equipped with a flexible range of measurement technologies including Absorbance (UV-Vis, including path length correction), Fluorescence intensity (including fluorescence resonance energy transfer (FRET)), Luminescence (direct and filtered, including bioluminescence resonance energy transfer (BRET)), Time-resolved fluorescence (TRF, including TR-FRET, homogeneous TRF (HTRF)).

Specification

Absorbance
Plate types6- to 384-well plates
Wavelength selectionDouble monochromators
Wavelength range200–1,000 nm
Light sourceXenon flash lamp
Read-out range0–6 Abs
Linear measurement range0–4 Abs (96-well plate) at 450 nm, ±2%; 
0–3 Abs (384-well plate) at 450 nm, ±2%
Accuracy0.003 Abs or ±2%, at 200–399 nm (0–2 Abs)
0.003 Abs or ±1%, at 400–1,000 nm (0–3 Abs)
PrecisionStandard deviation (SD) <0.001 Abs or coefficient of variation (CV) <0.5%, at 450 nm (0–3 Abs)
Fluorescence intensity
Plate types6- to 1,536-well plates
Wavelength selectionDouble excitation and emission monochromators
Excitation wavelength range200–1,000 nm
Emission wavelength range270–840 nm
Light sourceXenon flash lamp
SensitivityTop reading: <0.4 fmol fluorescein/well (black 384-well plate)
Bottom reading: <4 fmol fluorescein/well (clearbottom, black 384-square well plate)
Dynamic rangeTop reading: >6 decades
Bottom reading: >5.5 decades
Time-resolved fluorescence
Plate types6- to 1,536-well plates
Wavelength selectionFilters (spectral scanning with double excitation and emission monochromators)
Excitation wavelength rangeFixed to 334 nm (spectral scanning 200–840 nm)
Emission wavelength range400–700 nm (spectral scanning 270–840 nm)
Light sourceXenon flash lamp
Sensitivity<1 amol Eu/well (white, low-volume 384-well plate)
Dynamic range>6 decades
Luminescence
Plate types6- to 1,536-well plates (spectral scanning from 6- to 384-well plates)
Wavelength selectionDirect or filters (spectral scanning with double monochromators)
Wavelength range360–670 nm
Sensitivity<7 amol ATP/well (white 384-well plate)
Dynamic range>7 decades
AlphaScreen
Plate types6- to 1,536-well plates
Wavelength selectionFilters
Excitation wavelength rangeFixed to 680 nm
Emission wavelength range400–660 nm
Light sourceLED
Sensitivity<100 amol phosphotyrosine/well (white 384-well plate)
Dispensing
Plate types6- to 384-well plates
No. of dispensersNone, one, or two
Syringe size1 mL (standard), 5 mL (optional)
Dispensing volume2–5,000 µL, in 1 µL increments (1 mL syringe)
5–25,000 μL, in 5 µL increments (5 mL syringe)
Accuracy<1 μL with 50 μL (0.4 mm tip), <0.2 μL with 5 μL (0.25 mm tip)
Precision<1 μL with 50 μL (0.4 mm tip), <0.25 μL with 5 μL (0.25 mm tip)
Dead volumeReagent loss <100 μL, total tubing volume <800 μL
Incubator and shaker
Temperature rangeFrom ambient +4°C to 45°C
Shaking typeOrbital
Integrated gas module
CO2 concentration range0.1–15%
CO2 concentration stability±0.3% at 5% CO2
O2 concentration range1–21%
O2 concentration stability±0.3% at 1% O2
General features
Measurement modesEndpoint, kinetic, spectra, multipoint and kinetic spectra
Measurement speedReads a 96-well plate in 15 sec, a 384-well plate in 45 sec, and a 1,536-well plate in 135 sec (minimum times)
μDrop
Applications measurements of DNA, RNA, protein samples, or spectral scanning
Measurement rangedsDNA detection range from a few micrograms to a few milligrams per milliliter
Sample capacity and volume32 (µDrop Duo Plate) sample positions for quick and easy measurement of sample volumes down to 2 μL
FeatureCompatible with an eight-channel pipette for sample addition
SoftwareWith instrument control functions and advanced analytical functions

Brochure

Varioskan LUX_Brochure_EN