Single‑Quadrupole LC‑MS System
Equipment Introduction
Agilent 1290 Infinity Ⅲ‑Pro iQ -single quadrupole LC-MS system. Its LC module offers a maximum pressure tolerance of 1300 bar and low extra-column dead volume, enabling efficient and rapid separation. The integrated Pro iQ mass detector features fast scanning, a wide mass range and high sensitivity. With a compact footprint, it can be stacked with the UHPLC system and supports ESI and APCI ionization modes. Seamlessly interfaced with the matched LC platform and OpenLab CDS workstation, the instrument streamlines qualitative and quantitative workflows. Equipped with intelligent status alert and auto-tune functions, it delivers excellent system stability and easy maintenance. It is mainly used for qualitative and quantitative analysis of small molecules, impurity profiling, peptide molecular weight determination, and screening of environmental contaminants. Widely applied in drug development, metabolomics, proteomics and environmental analysis, the instrument is open for shared access by researchers.
Specification
| Agilent 1290 Infinity III UHPLC | |
| Quaternary Pump | Flow precision: RSD ≤ 0.07 % Flow rate range: 0.001 - 5 mL/min, minimum setting increment: ≤ 0.001 mL/min Operating pressure range: 0 1300 bar Gradient composition precision: RSD < 0.15 % Integrated degasser: 4 channels |
| Autosampler | Injection volume range: 0.1 - 20 μL, minimum increment: ≤ 0.1 μL Injection precision: RSD < 0.15 % Operating pressure range: 0 - 1300 bar Sample capacity: 2 drawers, 200 positions for 2 mL vials; microplate injection supported Sample compartment temperature control: 4 ℃ - 40 ℃ |
| Column oven | Temperature range: 4 - 110 °C Temperature control stability: ≤ ±0.03 °C Temperature control accuracy: ≤ ±0.5 °C |
| Diode Array Detector (DAD) | Number of photodiodes: 1024 signal channels: 8 Maximum sampling rate: 240 Hz Short term noise at 230 nm: ≤ ± 3 × 10⁻⁶ AU Drift at 230 nm: ≤ 0.5 × 10⁻³ AU/hr Wavelength range: 190 - 640 nm Programmable slit width: 1, 2, 4, 8 nm |
| Single Quadrupole Mass Spectrometer | |
| Ion source | ESI and APCI ion sources; cleaning and replacement of ion transfer unit without venting vacuum |
| Mass scan range | m/z 2 - 1600 |
| Maximum scan speed | 15,000 amu/s |
| Sensitivity | • ESI positive mode: Column infused 1 pg reserpine, S/N ≥200:1 at m/z 609.3 in SIM mode • APCI positive mode: Column infused 1 pg reserpine, S/N ≥70:1 at m/z 609.3 in SIM mode |
| Positive/negative polarity switching time | ≤15 ms |
| Dynamic range | ≥ 6×10⁶ |
| Minimum SIM dwell time | 1 ms |
| Resolution mode | 4 types, supporting 0.4 Da high resolution mode |
| Mass accuracy | ≤ ±0.1 Da |
| Mass stability | ≤0.1 Da /24 hrs |
Brochure
Agilent InfinityLab Pro iQ_Brochure_CN_1
Agilent InfinityLab Pro iQ_Brochure_CN_2