EEG/ERP System

Model:

actiCHamp Plus

Location:

124G, 1F, E4

Equipment Introduction

The Brain Products actiCHamp Plus system is a powerful tool for simultaneous EEG recording of up to 128 channels, with real-time electrode impedance measurement for optimal data quality. It features easy-to-prepare active electrodes and is highly integrated for multimodal recordings, including eye trackers, fNIRS, TMS, and more. This system is widely used in cognitive neuroscience research, brain-computer interface research, neural signal decoding, and clinical diagnosis of neurological disorders. Its applications span across various areas such as perception, attention, emotion, memory.

Specification

Sampling rates (combined with Recorder)128/96/64 channels + 8AUX: 25kHz, 32 channels + 8AUX: 100kHz
Number of channels32, 64, 96, 128
Number of AUX channels8
Maximum bandwidthDC-7500Hz
A/D conversion24-bit, 1 converter/channel
Input noise (<70Hz)2 μV p.p. (DC-35Hz)
Input voltage range+/- 400 mV (EEG), 0.3-4 V (AUX)
Common mode rejection (CMR) 100 dB
Trigger pulse input8-bit
Trigger pulse output8-bit
Electrodesactive electrodes, passive electrodes (dry electrodes, saline electrodes are also compatible)
Impedance measurementYes, 3-color LED display
Input impedance≥ 1 GΩ
Accuracy≤ 100 nV/bit
Acceptable skin contact impedance≥ 50 KΩ
Sensitivity (AC Mode)<0.51 nV/bit
Data recordingBrainVision Recorder
Data analysisBrainVision Analyzer
Stimulus presentationE-Prime