Protein Stability Analysis System

Model:

Prometheus Panta

Location:

124A, 1F, E4

Equipment Introduction

Prometheus enables you to measure thermal stability, aggregation, and particle sizing and dispersity in a single run – label-free, and without using lots of sample. Get trustworthy, high-resolution data on your protein's stability from Prometheus.

Specification

nanoDSFMeasurement parameters: ratio: Tonset, Tm, Ea, reversibility of unfolding 330 nm, 350 nm: Tm, Excitation: 280 nm;
Concentration range: 0.005 mg/mL – 250 mg/mL;
Inflection point precision @ 75 °C: ± 0.1 °C;
Initial ratio repeatability: 0.008.
DLSMeasurement parameters: Tscattering, Tsize, rH, PDI, kD;
Laser wavelength: 405 nm ± 5 nm;
Concentration range (DLS/SLS): 0.5 mg/mL for a 15 kDa protein, up to 40% w/v;
Size resolution: Down to 0.5 nm – 2 μm.
SLSMeasurement parameters: Tscattering, average scattering intensity, molecular weight, B22;
Measurement accuracy: ≤ 10% molecular weight.
BackreflectionMeasurement parameters: Tturbidity;
Measurement accuracy: Larger than 12.5 nm radius.

Brochure

Prometheus Panta_Brochure_EN